Description

The DC-series deviatoric triaxial cells are specifically designed for rock triaxial tests that do not require an external frame to apply axial loads. The cell includes a hydraulic built-in deviator chamber to apply the axial load on the specimen. The sample is fitted into a Teflon sleeve between hardened steel endcaps and the assembly is immersed in confining pressurized oil. Electrical feed-through connectors are embedded at the cell base to allow the use of in-vessel instrumentation. Top & bottom pore fluid ports at both upper and lower platens are an integral feature of the cell architecture. Different diameter platens can be supplied to perform custom sample geometry tests. If required, the cell can also be provided with a heating mantle.

For more information, please visit Floxlab's website: floxlab

Features

Manufacturer: Floxlab
Core diameter: up to 54.7 mm (model 1) and up to 100 mm (model 2)
Core length : twice the diameter
Compression axial load stress: 1,000 kN (model 1) and 2,500 kN (model 2)
Max confining pressure : 70/140 MPa
Max pore pressure : 70/140 MPa
Working temperature : ambient to 150° C
Wetted parts: Stainless steel
Power supply: 110-220VAC, 50/60Hz 

Benefits

* Built in deviatoric chamber, no external load frame required
* Models available for different specimen sizes